1
,Conditions of Use:1.1. The power supply meets the requirements of the instrument;
1.2. It should be grounded reliably when used.
2. Operating procedures:
Add water - stacking - sealing - set sterilization temperature and time - heating (discharge cold air) - sterilization - end
Add water: open the sterilizer cover, take out the sterilization barrel, and inject distilled water into the container to the water mark mark;
Stacking: Keep appropriate gaps between items when stacked;
Sealing: After the items are stacked, cover the sterilizer cover and tighten the hand wheel clockwise to make the cover close to the mouth, but not too tight, so as not to damage the sealing ring;
Set the sterilization temperature and time: adjust the setting button to set the required sterilization temperature and time;
Heating (discharging cold air): The power is turned on, and the power indicator and heating indicator are on. Before starting heating, rotate the bleed valve to the deflation position. After the cold air has fully escaped, close the bleed valve.
Sterilization: When the temperature and pressure rise to the preset sterilization temperature value, the heating lamp is off, the instrument enters the constant temperature control state for sterilization, and the timer automatically counts;
End: After the sterilization time reaches the set time, the buzzer will give an alarm and the sterilization will end. Turn off the power. When the pressure drops to zero, open the vent valve. After the steam is completely released, open the container and take out the sterilized items.
3, note:
3.1. Check whether the water level is at the water level mark before each sterilization;
3.2. When heating, the bleed valve must be moved to the deflation position to allow the cold air in the sterilizer to escape, otherwise the sterilization effect will be affected;
3.3. After the daily sterilization, the water in the sterilizer should be drained, the scale should be cleaned, and the new distilled water should be replaced to improve the sterilization quality and prolong the service life;
3.4. After each sterilization, the pressure must be reduced to zero, and the steam in the container can be completely removed before opening.
4. Maintenance:
4.1 Change distilled water at least once a week.
4.2 Clean the whole body once a month.